PMA2112 High Intensity UVA probe

Model PMA2112 Digital High Intensity UVA Probe

Solar Light’s Model PMA2112 Digital High Intensity UVA Probe is a useful tool for UV curing applications. It measures UVA radiation via a diffuser mounted at the end of its 18″ (45.7 cm) probe. The light is delivered to the sensor through a quartz light guide encapsulated in a metal envelope inside the probe chassis, making this probe suitable for high temperature applications up to 400°C.

Category:

Description

Solar Light’s Model PMA2112 Digital High Intensity UVA Probe is a useful tool for UV curing applications. It measures UVA radiation in the range from 320 to 400nm (with peak sensitivity at the 365nm mercury emission line) via a diffuser mounted at the end of its 18” (45.7 cm) probe. The light is delivered to the sensor through a quartz light guide encapsulated in a metal envelope inside the probe, making it suitable for high temperature applications up to 400°C. When used with our PMA2100 or PMA2200 Radiometers, the irradiance is displayed in W/cm2 or mW/cm2. The UVA sensor’s high dynamic range allows measurement of signals as strong as 10 W/cm2 and as weak as 0.1 mW/cm2. The probe is milled with a guide slot for repeatable positioning.

0.4” (1 cm)

This field can't be Empty

Vertical Downward, Vertical Upward, or Horizontal (for all models - please specify at order)

This field can't be Empty

Xenon Short Arc (For All Models)

This field can't be Empty

150W (For All Models)

This field can't be Empty

±5% (For All Models)

This field can't be Empty

A (IEC 60904-9 2007)<br> A (JIS C 8912)<br> A (ASTM E927 - 05)

This field can't be Empty

< ±2% rms

This field can't be Empty

B (IEC 60904-9 2007)<br> B (JIS C 8912)<br> B (ASTM E927 - 05)

This field can't be Empty

3” (7.6 cm)

This field can't be Empty

n/a

This field can't be Empty

Near Zero

This field can't be Empty

Factory Set Limit is 150 watts max

This field can't be Empty

32°F to 95°F / 0°C to 35°C

This field can't be Empty

-4°F to 185°F / -20°C to +85°C

This field can't be Empty

0 to 95% non-condensing

This field can't be Empty

Forced air

This field can't be Empty

EN61010-1 Laboratory, EN60335 Appliances, IEC60601-1 Medical

This field can't be Empty

EN55011 Emissions, IEC60601-1-2:2001, 2nd Rev 2 Medical, IEC61000-3-2 Harmonic, IEC61000-3-3 Flicker, IEC61000-4-2 ESD, IEC61000-4-3 Radiated, IEC61000-4-4 EFT, IEC61000-4-5 Surge, IEC61000-4-6 Conducted, IEC61000-4-11 Voltage Dip, IEC61000-4-8 Magnetic Field

This field can't be Empty

10 oz. (280 grams)

This field can't be Empty

<ul> <li>UV curing, printing and photolithography</li> <li>Monitoring of UV sources stability and lifetime</li> <li>Measurements in hazardous environment</li> </ul>

This field can't be Empty

<img src="https://solarlight.com/wp-content/uploads/PMA2112-log.jpg"><border="0">

This field can't be Empty

<img src="https://solarlight.com/wp-content/uploads/pma2112-vertical-cosine.jpg"><border="0">

This field can't be Empty

<img src="https://solarlight.com/wp-content/uploads/pma2112-horizontal-cosine.jpg"><border="0">

This field can't be Empty

10 [W/cm²] or 10,000 [mW/cm²]

This field can't be Empty

0.001 [W/cm²] or 0.1[mW/m²]

This field can't be Empty

<0.05%°C

This field can't be Empty

5ft (1.5m)

This field can't be Empty

1.6" (40.6 mm) [Sensor] <br> 0.375" (9.5 mm) [Probe]

This field can't be Empty

Brochure

UVA SENSOR PMA2112 UVA high intensity probe

Interfacing PMA detectors to a Campbell data logger

PMA Series capabilities and applications 

1 YEAR STANDARD WARRANTY

SOLAR LIGHT COMPANY warrants that the equipment has been carefully tested, inspected and left the Factory in proper working condition, free of visible defects.  SOLAR LIGHT COMPANY warrants the equipment to be free from defects in material and workmanship, under normal use and operation, for a period of one (1) year after the date of initial operation, or demonstration; or eighteen (18) months from the date of shipment from the Factory; or whichever occurs first. The warranty is limited to the free replacement of all-defective parts and free labor during the first year of warranty.  Defective parts replaced under this warranty shall become the property of SOLAR LIGHT COMPANY. The above stated Warranty is subject to the following conditions: Equipment being installed within the United States of America, Canada, or sold by authorized SOLAR LIGHT COMPANY representatives or their agents outside the continental United States. The equipment being installed is operated and maintained in accordance with SOLAR LIGHT COMPANY’s operating instructions, maintenance instructions and specifications. Utilization of only SOLAR LIGHT COMPANY’s authorized parts and components.  The utilization of a duly authorized representative for all work performed on SOLAR LIGHT COMPANY equipment during the warranty period.  Any variation(s) to the above warranty must be in writing and mutually acceptable by SOLAR LIGHT COMPANY and the customer. The warranty shall not apply to damages resulting from errors in installation, nor shall it apply to any equipment, which has been subject to damages, alterations or misuse by the purchaser.  Parts of the equipment, which would be generally considered as expendable service items during normal use, such as fuses, lamps, detector shields, etc., are not covered under the warranty. THERE ARE NO WARRANTIES, WHICH EXTEND BEYOND THIS DESCRIPTION, AND EXPRESSLY NO IMPLIED WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.  IN NO EVENT SHALL SOLAR LIGHT COMPANY BE LIABLE FOR SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR ANTICIPATED LOSS OF PROFITS SUFFERED BY THE PURCHASER. Extended Warranty Plans are also available.